For Professional Users OnlyEnglish BB0339-MJune 2012 User GuideType 8808Biplane Transducer
10 June 20128808 User Guide (BB0339-M)Figure 4. Assembly of the water standoff facility.4To remove all air from the rubber sheath, hold the transduce
8808 User Guide (BB0339-M) Puncture and Biopsy Facilities 11 Puncture and Biopsy FacilitiesPuncture and biopsy are possible with 8808. The appropriat
12 June 20128808 User Guide (BB0339-M)The transducer is prepared for transrectal puncture or biopsy by pulling a sterile transducer cover over the tr
8808 User Guide (BB0339-M) Puncture and Biopsy Facilities 13 Please refer to Care, Cleaning and Safety for an example of how to open a sterile-packed
14 June 20128808 User Guide (BB0339-M)For Transperineal PunctureThe UA1232 metal puncture attachment (see Figure 8) is designed for transperineal pun
8808 User Guide (BB0339-M) Performing Puncture and Biopsy 15 Performing Puncture and BiopsyIf not sterilized, cover the transducer with a sterile tra
16 June 20128808 User Guide (BB0339-M)To remove the puncture line from the image, refer to the applicable system user guide for instructions.Figure 1
8808 User Guide (BB0339-M) 3D Imaging 17 3D Imaging3D imaging with the 8808 may be carried out in either of two ways:•Freehand – where the transducer
18 June 20128808 User Guide (BB0339-M)
BK MEDICALMileparken 342730 HerlevDenmarkTel.:+45 4452 8100 / Fax:+45 4452 8199 www.bkmed.comEmail: [email protected] serial number of a BK Medical pro
LEGAL MANUFACTURER : BK Me dic al ApS, Mileparken 34, 2730 Herlev, Denma rk. Tel.: +45 44528100 Fax : +45 44528199 E mail: info@bkme d.
3 ContentsIntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Indications for
8808 User Guide (BB0339-M) Introduction 5 IntroductionThis is the user guide for Biplane Transducer Type 8808 and must be used together with Care, Cl
6 June 20128808 User Guide (BB0339-M)Simultaneous Biplane Imaging with 2202 system1The transducer can transmit transverse (T) and sagittal (S) images
8808 User Guide (BB0339-M) Cleaning and Disinfection 7 Service and RepairCaring for the TransducerThe transducer may be damaged during use or process
8 June 20128808 User Guide (BB0339-M)Connecting the TransducerThe transducer is connected to the system using the array Transducer Socket on the syst
8808 User Guide (BB0339-M) Imaging with Type 8808 9 Changing OrientationTo change the orientation of the image on the monitor, refer to the applicabl